国产精品99久久久久久www_乱小说录目伦合集最新章节_国产区小视频_一级毛片aa免费观看

Products
YG469 mask bacterial filtration efficiency tester

product description

The Bacterial Filtration Efficiency (BFE) detector for masks is composed of an aerosol generation system, aerosol chamber and aerosol transmission device, negative pressure operating system, 28.3L/min sampler, etc. The entire detector is uniformly controlled by a control program. The aerosol generation system includes: peristaltic pump, aerosol generator, compressed air and purification device. The aerosol chamber and aerosol transmission device include: aerosol chamber, high-efficiency air filter, suction negative pressure fan, etc. The console adopts a PLC control system to coordinate and control the work and monitoring of the aerosol generation system, transmission system, negative pressure operating system, and sampling system, and displays the working status in real time, and the entire measurement work is completed automatically. The bacterial filtration efficiency tester adopts the advanced design ideas and manufacturing technology in the same industry in the world. The main performance indicators of this set of instruments meet the standard YY/T 0969 disposable medical masks and YY 0469 medical surgical standards and refer to the relevant standards of ASTM F2100 and EN14683. On this basis, improvements have been made, and a two-channel sampling system is designed, sampling at the same time. Its excellent performance is especially suitable for the performance test of bacterial filtration efficiency (BFE) by metrological inspection departments, scientific research institutes, medical mask manufacturers and other related departments.

Guideline

YY/T 0969-2013, YY 0469-2011, ASTM F 2100, ASTM F 2101, EN 14683

technical parameter

 The main parametersParameter range

 Vacuum pumpOil-free rotary vane vacuum pump, maximum flow rate up to 57LPM, equipped with silencer, stable air flowAerosol generator flow control The flow control of the aerosol generator adopts mass flow control, which is accurate;

I Pneumatic sampling flowAnderson sampling flow rate 28.3L/min, resolution: 0.1L/min (user can calibrate)

II Pneumatic sampling flowAnderson Sampling Flow 28.3L/min Resolution: 0.1L/min (User can calibrate)

I and II samplingEquipped with mass flow controller

I Air pressure 0-20kPa, resolution: 0.01kpa

II Air pressure: 0-20kPa, resolution: 0.01kpa

Spray flow meter pressure: 0300kPa, resolution: 0.01kpa

Negative pressure of aerosol chamber: 0120Pa, resolution: 0.01kpa

Operating temperature: 050

Cabinet negative pressure: 0120Pa

Data storage capacity: Data can be downloaded via U disk

Mass flow controller: 020L/min Resolution: 0.1L/min

Peristaltic pump flow: (0.013.0)mL/min Resolution: 0.1L/min

High efficiency air filter characteristics: The filtration efficiency for particles above 0.3um is 99.99%

Mass median diameter of aerosol generator: Average particle diameter (3.0±0.3)μm, geometric standard deviation 1.5

Dual-channel 6-stage Anderson sampler capture

Particle sizeClass >7μm, Class (4.77)μm, Class (3.34.7)μm, Grade (2.13.3)μm, Grade (1.12.1)μm, Grade VI (0.61.1)μm

Aerosol chamber specifications600mm×80mm×5mm[L*D*T]

Positive quality control Total number of particles in the sampler(2200±500)cfu